When the temperature goes below 32 °F (0 °C), please observe the following steps while your centrifuge is at standstill. After shutdown, there is still some residual water left over in the machine. This water can freeze up. The water must be cleared or removed from:
In most cases, you can drain trapped water by opening the provided drain valves and setting the ball valves at 45°. On occasion, coolers (like the one on the control cabinet) need the assistance of air. Water can also trap behind solenoids and the centrifuge may need to be energized to completely perform these tasks.